Quotation |
| Ordering information |
See HOW TO ORDER below the CML technical specification
CML - SPECIFICATION
|
|
| Technology Laser package Sensitivity Accuracy Calibration Analysis range Sample Volume Operation Viscosity range Operating temperature Electrical requirement Fluid compatibility Data storage Dimensions Overall Built-in optional product |
Automatic Optical Particle Analyser Twin Laser and Twin Optical Diode Detectors > 4, 6,14,21,25,38,50, 68 µm (c), micron range to revised ISO 4406 Standard > 3% typical Each unit is individually calibrated with ISO Medium test Dust (MTD) based on ISO11171:1999 on equipment certified by I.F.T.S. ISO Code 8 to 24 to ISO 4406. NAS 1638 Class 2 to 12 AS4059 Size Codes A:000 to 12, B00 to 12, C00 to 12, D:2 to 12, E 4 to 12, F: 7-12 15ml (normal) 30ml (dynamic) 24ml (triple) 15ml (continuous) 8ml (short) Max. system pressure: 400 bar. Min. working pressure: 2 bar. To 400 centistokes +5 to +80°C 12 to 24V DC, 1 amp max. Mineral oil & petroleum based (consult MP Filtri for other fluids) Approximately 600 tests Height 152mm Width 295mm Length 340mm Weight 4.75 kg Moisture and temperature measurement included on the CML2-W model* The phosphate ester product with moisture sensor is not currently available. (Please contact MP FILTRI) |
< select buttons for
further information